홈MCPSkills워크벤치OpenChem
홈워크벤치알림마이페이지
llms.txt · skill.md · MCP 가이드개인정보처리방침 · 이용약관 · 지원v1.8.5
  1. 홈
  2. /반응

Acetyl chloride → Sec-Butyl acetate

HRID 579422
인쇄

반응식

[CH:1](N([CH:7]([CH3:9])[CH3:8])CC)(C)[CH3:2].CN(C1C=CC=CN=1)C.C(Cl)(=O)C.[C:23]([O:26][C:27](=[O:29])[CH3:28])(=[O:25])[CH3:24]>C(Cl)Cl.C(OCCCC)(=O)C.C(OCC)(=O)C>[C:23]([O:26][CH:7]([CH3:8])[CH3:9])(=[O:25])[CH3:24].[C:27]([O:26][CH:23]([CH2:1][CH3:2])[CH3:24])(=[O:29])[CH3:28]
참가 물질실험 절차후처리출처 및 근거반응 노트

참가 물질

9

반응물 4

HCID 6367Acetyl chloride2회HCID 7918Acetic AnhydrideHCID 218852-Dimethylaminopyridine2회HCID 81531Diisopropylethylamine2회

생성물 2

HCID 7758Sec-Butyl acetateHCID 7915Isopropyl Acetate

시약·촉매·용매 3

HCID 6344
Dichloromethane
용매
2회
HCID 8857Ethyl Acetate용매
HCID 31272Butyl Acetate용매

실험 절차

Samples were analyzed on a Agilent 6890 5973 GCMS system equipped with a JW1 DB624 column with dimensions of 30 m×250μ×1.4μ. The method ran at 1 ml/min flow, with oven temperature at 40° C. for the first two minutes followed by temperature ramp at 10° C./min to a temperature of 240° C. which was held for 10 minutes. The solvent delay was set at 5 minutes. Chemical identities of compounds were confirmed by mass spectroscopic qualitative analysis on GCMS against a NIST 2011 library as well as by comparison of retention time against commercial standards. Analysis of samples with volatile alcohols was done on a Gowmac GC system using a Hayesep Q column with He and nitrogen carrier gases. Column temperature was 180° C., detector temp. was 150° C., injector temperature was 150° C., sample valve temperature was 105° C. detector current was 107 mA and injection volume was 1 microliter. Gas samples were analyzed on Varian Micro-GC instruments. Additionally, some product mixtures were analyzed for presence of lower alcohols by derivatizing the alcohol mixture by dissolving in methylene chloride, adding excess of diisopropylethylamine (DIEA), dimethylamino pyridine (DMAP) and acetyl chloride or acetic anhydride. After standing for 30 minutes the reaction mixture was analyzed by GCMS. An experiment for lower molecular weight alcohol analysis was done as follows: The liquid product (30 μl) of the reaction was dissolved in 1 ml of methylene chloride, DIEA was added with a 10 μl syringe in increments until the medium was basic, 2 mg of DMAP was added followed by acetyl chloride. The solution was allowed to stand for about a half hour until the alcohols were esterified to their acetyl derivatives. The resulting mixture was analyzed on the GCMS giving a mixture with ethyl acetate, isopropyl acetate, butyl acetate and sec-butyl acetate. These values represented the approximate percentages of ethyl, isopropyl, butyl alcohols in a sample.

후처리

  1. customequipped with a JW1 DB624 column with dimensions of 30 m×250μ×1.4μ
  2. customThe method ran at 1 ml/min flow, with oven
  3. customat 40° C.
  4. customfor the first two minutes
  5. customfollowed by temperature ramp at 10° C./min to a temperature of 240° C. which
  6. waitThe solvent delay was set at 5 minutes
  7. customwas 180° C.
  8. customwas 150° C.
  9. customwas 150° C.
  10. customsample valve temperature was 105° C.
  11. waitAfter standing for 30 minutes

출처 및 근거

ORD 기록
ord-1bac650712ff4de5840f843f82f029df
출처 데이터셋
uspto-grants-2015_05
DOI
10.6084/m9.figshare.5104873.v1
특허 번호
US09040757B2

반응 노트

이 페이지에서참가 물질실험 절차후처리출처 및 근거반응 노트